Temperature measurement

When does the temperature get measured and what are the recommended temperatures?

Written by Assaf Shafran

After 20 injections, the system starts measuring the temperature.

From X.4.12 Firmware version-
The temperature presented is the actual temperature detected.

The temperature is measured only during vaccine flow.
Meaning while injecting, priming, calibrating, and cleaning & disinfecting.
When the vaccine stops flowing, the last detected temperature will be displayed.

The reading accuracy of the temperature sensor can be +/-1 degrees.

In the previous Firmware versions-

The temperature displayed is the average of the last 10 measurements performed once a second.

Vaccine temperature

Set by the user for each vaccine separately in the Mobile App or in the Web App.

Recommended vaccine temperature: 25 - 35°C

Cleaning & disinfection temperature

The water needs to be between 50°-60°C / 122°-140°F.

The device will give a Red light indication on the Control unit and the Hand unit if it is above 65°C / 149°F.

The system will give an Extremely high alert - a Red light indication on the Control and Hand units if it is above 75°C / 167°F.