SoundPro

How to set up TSI Quest SoundPro sensors

Below you'll see the step by step process for setting up the SoundPro series of sensors as well as tips for troubleshooting and understanding model specific features. Check out guides on how to physically set up your Thiamis, and how to add a sensor in Environet.

Quest SoundPRO Special Requirements

The Quest SoundPro requires the serial connector cable (TSI part number 053-729) which plugs into the auxiliary port and converts it to a RS-232 serial port.

Note: The images in this article show an older Thiamis, but the setup is the same for modern Thiamis devices. The only difference is that all USB-A ports have been replaced with USB-C ports on the modern devices.

Setup Instructions:

  1. Connect power to the Thiamis, there are two options:

    • Connect a 12V power supply to the 3 pole Euroblock connector

    • Connect a 5V power supply to either of the USB ports

  2. Plug the adapter cable into the SoundPro's auxiliary serial port, on the bottom of the sensor.

  3. Connect the other end of the cable directly to the RS-232 port on the Thiamis.

Note: When using a splitter cable, note which sensor is connected to each letter port for configuring in Environet when Adding Sensors


Errors

If the SoundPro is not reporting, check the Thiamis configuration page for any listed Errors. Errors can be found in Sensor Configuration section of the Thiamis Configuration Page, and will be highlighted in RED.

Timeout Error

The SoundPro has not responded to the requests made by the Thiamis and has timed out. It will continue to attempt communication but you can run through the general Troubleshooting Sensors steps to address common problems.


Tips on Specific Models:

SoundPro SE and SoundPro DL

Only A-weighted Lav, Leq, Lpk, Lmin, Lmax are supported.

Settings

Ensure data output is enabled by confirming that your settings are as follows:

START menu > SETUP > COMM-SET:

  • USB: QSP/Serial

  • RS-232: Serial

  • Baud Rate 4800

Console Controls

When a SoundPRO sensor is connected to a Thiamis, configuration options become available on the Thiamis Configuration Page, more about sensor configuration can be found in this article: Adding and Configuring sensors.

The sensor configuration options for SoundPro sensors are:

  • Measurement Mode - Select between SLM, 1/1, and 1/3

  • Filters

  • Session Time - toggle on or off


List of Supported Models: